dc.description.abstract |
Patients with hemoglobin (Hb) variants may produce false hemoglobin A1c (HbA1c) measurement leading to misdiagnosis or mistreatment. This study aimed to evaluate the effects of Hb E on five different HbA1c analytical principles. HbA1c levels of 160 EDTA blood samples including Hb type A2A (N=51), heterozygous Hb E (N=104), and homozygous Hb E (N=5) were analyzed by immunoturbidimetric method, HPLC, LPLC, enzymatic assay, and boronate affinity chromatography. The strong positive correlations of HbA1c levels measured by all assays were observed (r=0.797 to 0.962, p<0.001). The HbA1c results analyzed by immunoassay and enzymatic assay were significant higher than boronate affinity chromatography (p<0.05), in Hb type A2A and heterozygous Hb E samples. However, these differenced between assays in each hemoglobin typing were comparable. The results by using HPLC were under reportable range in homozygous Hb E group. The mean differences between HbA1c results analyzed by using LPLC and other 4 testes methods were significantly higher in heterozygous Hb E samples with Hb F>2%, compared to samples with Hb F <=2% (p<0.05). This work showed a good comparability of HbA1c assays, although, some methods were influenced by high levels of Hb E and Hb F. Therefore, Hb E and Hb F are interfering factors that laboratories must be aware when reporting results if its presence is suspected. |
en |
dc.language.iso |
en_US |
en |
dc.rights |
มหาวิทยาลัยหัวเฉียวเฉลิมพระเกียรติ |
en |
dc.subject |
ฮีโมโกลบินเอวันซี |
en |
dc.subject |
Hemoglobin A1c protein, human |
en |
dc.subject |
ฮีโมโกลบินอี |
en |
dc.subject |
Hemoglobin E |
en |
dc.subject |
ไฮเปอร์ฟอร์แมนซ์ลิควิดโครมาโตกราฟี |
en |
dc.subject |
High performance liquid chromatography |
en |
dc.subject |
ลิควิดโครมาโตกราฟี |
en |
dc.subject |
Liquid chromatography |
en |
dc.title |
Effect of Hemoglobin E on Measurement of Hemoglobin A1c by Fire Different Methods |
en |
dc.type |
Proceeding Document |
en |